Pharmaceutical


Residue on Ignition <281>
Product Number PEP-006
Price:   $81.60
Availability:  Available

 Purchase Product


Select Study Option

Quantity   
NOTICE:  Product price and availability vary by location. Select your location to ensure price and availability.


PEP-006 Residue on Ignition (ROI) - Kit



PEP006-1EA Residue on Ignition (ROI) - PT


General Instructions

Ignite suitable crucibles at 600 °C ± 50 °C for 30 minutes, cool in a desiccator and weigh accurately.

Weigh accurately approximately 1 g of the test sample iinto the crucible.

Moisten with about 1 mL of sulfuric acid and heat gently by an appropriate method until the sample is thoroughly charred.

Cool then repeat with another portion of sulfuric acid (about 1 mL) until white fumes are no longer evolved.

Then ignite at 600 °C ± 50 °C until the residue is completely incinerated.

Calculate the % of residue.

**NOTE** Smaller quantities of test sample may be used.


Report

% Residue to two significant figures


Sample B Instructions

Expected Range: 0.01 to 15%



PEP-006-A Residue on Ignition (ROI) - Sample A


General Instructions

Ignite suitable crucibles at 600 °C ± 50 °C for 30 minutes, cool in a desiccator and weigh accurately.

Weigh accurately approximately 1 g of the test sample iinto the crucible.

Moisten with about 1 mL of sulfuric acid and heat gently by an appropriate method until the sample is thoroughly charred.

Cool then repeat with another portion of sulfuric acid (about 1 mL) until white fumes are no longer evolved.

Then ignite at 600 °C ± 50 °C until the residue is completely incinerated.

Calculate the % of residue.

**NOTE** Smaller quantities of test sample may be used.


Report

% Residue to two significant figures


Sample A Instructions

Expected Range: 0.01 to 15%



PEP-006-C Residue on Ignition (ROI) - Sample C


General Instructions

Ignite suitable crucibles at 600 °C ± 50 °C for 30 minutes, cool in a desiccator and weigh accurately.

Weigh accurately approximately 1 g of the test sample iinto the crucible.

Moisten with about 1 mL of sulfuric acid and heat gently by an appropriate method until the sample is thoroughly charred.

Cool then repeat with another portion of sulfuric acid (about 1 mL) until white fumes are no longer evolved.

Then ignite at 600 °C ± 50 °C until the residue is completely incinerated.

Calculate the % of residue.

**NOTE** Smaller quantities of test sample may be used.


Report

% Residue to two significant figures


Sample C Instructions

Expected Range: 0.01 to 15%



strammel@rt-corp.com